This port style is found exclusively on vessels manufactured by B. Braun Biotech.
Part Number | Size and Material | Compatible Plugs | Compatible Housings | Installation | Certs |
Model 210 (AF-210-60-L040) | 25 mm Angled Threaded Port316L Stainless Steel
Side Mount |
Model 305 | Model 326 Model 336 |
Installation Guide | A Material Technical Report is sent with each order |
The electrode housing is designed to hold and protect the pH electrode while inserting it into the bioprocess vessel. There are various types of side entry ports and different groups of housings to fit each port type. The size and make of the port must first be identified in order to narrow the selection process.
Standard 25mm Side Port
Found on nearly all vessels other than those manufactured by B. Braun Biotech. The port bore is 25 mm i.d. and the housing o-ring seals to the inside of the port. The housing is then secured to the port by a threaded retainer ring. The port is installed at a 15° angle for better electrode performance. See the cutaway drawing to the right for typical installation dimensions.
B. Braun Biotech 25 mm Safety Port
Found exclusively on vessels manufactured by B. Braun Biotech. This style of port is longer than the standard port and needs a special housing to fit correctly. (Note: the port opening on newer tanks is 30 mm i.d. and the port narrows down to 25 mm i.d. at the critical point where the housing o-ring seals to the inside wall of the port.) Again, the port is installed at a 15° angle for better electrode performance. See the cutaway drawing to the right for typical installation dimensions.